Argon Dewar Relief Set Pressure Modifications
Author(s) -
Jianzhou Wu
Publication year - 1991
Publication title -
osti oai (u.s. department of energy office of scientific and technical information)
Language(s) - English
Resource type - Reports
DOI - 10.2172/1031772
Subject(s) - cryostat , argon , set (abstract data type) , relief valve , value (mathematics) , high pressure , chemistry , materials science , thermodynamics , mechanics , mechanical engineering , engineering , mathematics , computer science , physics , statistics , condensed matter physics , programming language , superconductivity , organic chemistry
This engineering note documents the calculations of Kelly Dixon, used to determine the maximum allowable set pressure for the argon dewar low relief valve, tag number PSV620A, Anderson Greenwood Type 526J. The original setting was 16 psig. This value was chosen in order to protect against cryostat overpressurization by the source dewar (see D0 Engineering Note 115), however, the following calculations will show that the set pressure can be raised to approximately 18.5 psig, which would result in a faster filling of the cryostat, along with a higher level of liquid argon. Three other engineering notes were revised to reflect the change in set pressure according to this note. They are notes 115, 219, and 263
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom