Low-Cost System for Real-Time Monitoring of Luciferase Gene Expression
Author(s) -
Paul C. Gailey,
Estelle Miller,
Guy D. Griffin
Publication year - 1997
Publication title -
biotechniques
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.617
H-Index - 131
eISSN - 1940-9818
pISSN - 0736-6205
DOI - 10.2144/97223rr04
Subject(s) - luciferase , gene expression , biology , gene , computational biology , microbiology and biotechnology , genetics , chemistry , transfection
In some mammalian cells transfected with luciferase reporter genes, the luciferase/luciferin reaction in a cell monolayer produces a very small light flux. While the low light levels are often measurable with single-photon counting cameras, these devices are expensive and may require long averaging times to acquire an image. We describe an approach for real-time monitoring of light produced by luciferase gene expression in intact, cultured cells using readily available and relatively inexpensive components. The system uses a single-photon counting photomultiplier tube with built-in high voltage supply and photon counting circuitry to rapidly measure average light output from growing cells in a 35 mm culture dish. The fast, accurate and highly sensitive response of the system makes it useful for studying the dynamics of gene expression over time periods ranging from minutes to days.