The LightCycler TM : A Microvolume Multisample Fluorimeter with Rapid Temperature Control
Author(s) -
Carl T. Wittwer,
Kirk M. Ririe,
Richard Andrew,
Derek David,
R.A. Gundry,
Ulysses J. Balis
Publication year - 1997
Publication title -
biotechniques
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.617
H-Index - 131
eISSN - 1940-9818
pISSN - 0736-6205
DOI - 10.2144/97221pf02
Subject(s) - fluorometer , fluorescence , cuvette , analytical chemistry (journal) , photomultiplier , materials science , analyte , chemistry , optics , chromatography , physics , detector
Experimental and commercial microvolume fluorimeters with rapid temperature control are described. Fluorescence optics adopted from flow cytometry were used to interrogate 1-10-microL samples in glass capillaries. Homogeneous temperature control and rapid change of sample temperatures (10 degrees C/s) were obtained by a circulating air vortex. A prototype 2-color, 32-sample version was constructed with a xenon arc for excitation, separate excitation and emission paths, and photomultiplier tubes for detection. The commercial LightCycler, a 3-color, 24-sample instrument, uses a blue light-emitting diode for excitation, paraxial epi-illumination through the capillary tip and photodiodes for detection. Applications include analyte quantification and nucleic acid melting curves with fluorescent dyes, enzyme assays with fluorescent substrates and techniques that use fluorescence resonance energy transfer. Microvolume capability allows analysis of very small or expensive samples. As an example of one application, rapid cycle DNA amplification was continuously monitored by three different fluorescence techniques, Which included using the double-stranded DNA dye SYBR Green I, a dual-labeled 5'-exonuclease hydrolysis probe, and adjacent fluorescein and Cy5z-labeled hybridization probes. Complete amplification and analysis requires only 10-15 min.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom