Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.
Author(s) -
Stephen Havas,
Robert H. Bishop,
Lauren Koumjian,
Joel I. Reisman,
S Wozenski
Publication year - 1992
Publication title -
american journal of public health
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.284
H-Index - 264
eISSN - 1541-0048
pISSN - 0090-0036
DOI - 10.2105/ajph.82.3.458
Subject(s) - fingerstick , medicine , medical screening , endocrinology , diabetes mellitus
The precision, accuracy, and durability of the Reflotron were evaluated by the Massachusetts Model Systems for Blood Cholesterol Screening Program. Screenings were conducted in diverse community settings over 16 months. Fingerstick samples from 10,428 individuals were tested. None of the four analyzers met the 1992 standards for precision, although two met the 1992 standards for accuracy. More than 40% of Reflotron values differed from the reference laboratory values by upwards of 5%. More than 16% of individuals were misclassified in terms of their risk category. All four instruments malfunctioned during the project.