z-logo
open-access-imgOpen Access
MR-Safe Pneumatic Rotation Stepping Actuator
Author(s) -
Hiroyuki Sajima,
Hiroki Kamiuchi,
Kenta Kuwana,
Takeyoshi Dohi,
Ken Masamune
Publication year - 2012
Publication title -
journal of robotics and mechatronics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.257
H-Index - 19
eISSN - 1883-8049
pISSN - 0915-3942
DOI - 10.20965/jrm.2012.p0820
Subject(s) - actuator , rotation (mathematics) , puncturing , rotary actuator , torque , computer science , robot , robotics , distortion (music) , magnetic resonance imaging , angular velocity , pneumatic actuator , rotational speed , simulation , artificial intelligence , biomedical engineering , mechanical engineering , engineering , physics , medicine , radiology , telecommunications , amplifier , computer network , bandwidth (computing) , quantum mechanics , thermodynamics
Magnetic Resonance Imaging (MRI) is widely used not only for diagnosis but also for surgical navigation, etc. Surgeons and surgery-support systems can conduct surgeries less invasively because they derive accurate positional relationships between surgical instruments and anatomical regions of interest from preoperative or intraoperative MRI images. Many surgical robots intended for use within MRI gantries have been developed. Some of them use pneumatic actuators that are difficult to control accurately or that have many components, however, causing high fabrication cost, low durability and low sterilizability. To solve these problems, we have developed a φ30-mm pneumatic rotation stepping actuator. The actuator consists of three Direct Acting gears (D.A. gears) and a Rotation gear (R gear). When all of the D.A. gears are pushed sequentially up by compressed air, the R gear rotates because it engages all of them. In a fundamental performance experiment, the maximum angular error was 2.1° and maximum torque was approximately 150 mNm using 0.6 MPaG. Additionally, in an MR safety evaluation, the actuator was not found to cause any distortion or artifacts in MRI images. The actuator can therefore be applied to MR-safe positioning or puncturing robots.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom