z-logo
open-access-imgOpen Access
Prosthetic Fracture of a Cemented Exeter Femoral Stem (Case Report)
Author(s) -
Michael Facek,
Yasser Khatib,
Eric Swarts
Publication year - 2016
Publication title -
reconstructive review
Language(s) - English
Resource type - Journals
eISSN - 2331-2270
pISSN - 2331-2262
DOI - 10.15438/rr.6.4.159
Subject(s) - prosthesis , dimple , orthodontics , context (archaeology) , fracture (geology) , femoral neck , materials science , medicine , surgery , composite material , geology , osteoporosis , paleontology , endocrinology
We present a single case of fracture of an Exeter femoral prosthesis at the neck, occurring after a fall from standing height, in a lean 70yr lady at 6 years post implantation. The fracture propagated from the insertion dimple on the superior aspect of the prosthesis shoulder. Materials analysis suggested variance in composition of the alloy, particularly with grain size heterogeneity. Whilst Exeter femoral prosthesis fracture is extremely rare, when it does occur the literature suggests it is often in the context of excessive mechanical stresses (obesity, high offset, falls). However, this case may represent a failure of materials rather than mechanical stresses alone.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom