Performance of HemosIL ReadiPlasTin, a Liquid Reagent for Prothrombin Time Measurement
Author(s) -
Saeam Shin,
Yunjung Jung,
Wonkeun Song,
Min-Jeong Park
Publication year - 2019
Publication title -
imsang byeong'ri wa jeongdo gwanri/journal of laboratory medicine and quality assurance
Language(s) - English
Resource type - Journals
eISSN - 2288-7261
pISSN - 1225-097X
DOI - 10.15263/jlmqa.2019.41.1.9
Subject(s) - prothrombin time , reagent , medicine , chromatography , chemistry , organic chemistry
Saeam Shin, Yunjung Jung, Wonkeun Song, and Min-Jeong Park Department of Laboratory Medicine, Hallym University Kangnam Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea Background: Prothrombin time (PT) measurement is an important test for screening blood coagulation disorders and monitoring anticoagulant therapy. In this study, we evaluated the analytical performance of HemosIL ReadiPlasTin (Instrumentation Laboratory, USA), a liquid reagent for PT measurement. Methods: The precision of HemosIL ReadiPlasTin was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) EP5-A3 guidelines. Further, comparison with HemosIL RecombiPlasTin 2G (Instrumentation Laboratory, USA) was made according to the CLSI EP9-A3 guidelines. The reference intervals were established according to the CLSI C28-A3 guidelines. Results: The coefficient of variation values for repeatability and total imprecision at two levels of control materials were lower than 1.1% and 3.4%, respectively. The performance of HemosIL ReadiPlasTin was comparable to that of HemosIL RecombiPlasTin 2G, with a high correlation (r=0.996). The reference interval for normal subjects was 10.4–13.3 seconds. Conclusions: HemosIL ReadiPlasTin showed an acceptable degree of imprecision and its performance showed high correlation with that of a conventional reagent. Therefore, it is expected to be useful for PT measurement in clinical laboratories. (J Lab Med Qual Assur 2019;41:9-12)

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom