[Optimum hepatic parenchymal dissection to prevent bile leak: a comparative study using electrosurgical and stapling devices in swine].
Author(s) -
Tetsuo Ikeda,
Tomohiko Akahoshi,
Hirofumi Kawanaka,
Hideaki Uchiyama,
Yo-ichi Yamashita,
Masaru Morita,
Eiji Oki,
Hiroshi Saeki,
Keishi Sugimachi,
Toru Ikegami,
Tomoharu Yoshizumi,
Yuji Soejima,
Ken Shirabe,
Koshi Mimori,
Masayuki Watanabe,
Makoto Hashizume,
Yoshihiko Maehara
Publication year - 2013
Publication title -
fukuoka igaku zasshi = hukuoka acta medica
Language(s) - English
DOI - 10.15017/1440929
BACKGROUNDBile leakage is a serious complication of liver resection, and its treatment is very time-consuming. In open liver resection, Glisson's sheaths are usually disconnected by ligation to the extent possible during the parenchyma dissection. However, in laparoscopic surgery, the ligation, suture, and hemostasis are more difficult than in open surgery. For this reason, in laparoscopic liver resection, liver parenchyma dissection is generally accomplished using electrosurgical or stapling devices.PURPOSEThe purpose of this study was to verify the authenticity of electrosurgical devices attached an automatic irrigation function (AI) and stapling devices for laparoscopic liver parenchymal dissection.METHODSFour devices were used for liver parenchymal dissection in laparoscopic hepatic wedge resection, in pigs: monopolar high-frequency electric cautery attached AI (MCI) (n = 6), bipolar high-frequency electric cautery attached AI (BCI) (n = 6), bipolar tissue sealing system (LigaSure) attached AI (BSI) and an endoscopic stapling device (ECHELON FLEX ENDOPATH) (ES). In each group, burst pressures were tested using an electronic manometer, paying special attention to the location (s) of the first disruption (s). The dissected tissues were examined histologically.RESULTSPressures used in electrosurgical devices attach AI were significantly higher compared to pressures used in a ES (P < 0.001). While thermal denaturation of the liver parenchyma occurred at approximately 2-3 mm of depth when bipolar high-frequency electric cautery was used for dissection, it reached up to more than 10 mm with monopolar high-frequency electric cautery. All of the first disruption points of stapling were at stapling line.CONCLUSIONSElectrosurgical devices with an automatic irrigation function are useful devices to dissect the liver parenchyma.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom