Analytical and Clinical Evaluation of the NOWDiagnostics ADEXUSDx Human Chorionic Gonadotropin Test Using Whole Blood
Author(s) -
Robert D Nerenz,
Jennifer RC Bell,
Nancy Montes de,
Joann Short,
Theresa Mims,
Patrick A Cleeton,
J.D. Moore,
R. Humphries
Publication year - 2016
Publication title -
the journal of applied laboratory medicine
Language(s) - English
Resource type - Journals
eISSN - 2576-9456
pISSN - 2475-7241
DOI - 10.1373/jalm.2016.020297
Subject(s) - human chorionic gonadotropin , hematocrit , medicine , concordance , fingerstick , urine , reproducibility , pregnancy , pregnancy test , gynecology , andrology , whole blood , point of care testing , obstetrics , endocrinology , chemistry , chromatography , hormone , immunology , biology , genetics , diabetes mellitus
BACKGROUNDPoint-of-care (POC) urine qualitative human chorionic gonadotropin (hCG) devices are used to rapidly assess pregnancy status, but many of these devices are susceptible to false-negative results caused by increased concentrations of hCG β core fragment (hCGβcf) that does not contain hCGβcf.METHODSPurified hCG was added to hCG-negative heparinized whole blood to generate samples with known hCG concentrations, and the resulting samples were used to evaluate device sensitivity, low-end reproducibility, high-dose hook effect, intermediate range performance, acceptable sample volume, acceptable hematocrit range, and lot-to-lot variation. Device performance was also prospectively evaluated in 40 pregnant and 40 nonpregnant women aged 18-44 years in a hospital-based clinic or an academic hospital emergency department.RESULTSAll device observations were positive using a whole blood sample containing a plasma hCG concentration of 2.2 × 106 IU/L, and all device observations were positive from18 IU/L to 1.2 × 103 IU/L and from 2.5 × 104 IU/L to 2.2 × 106 IU/L. Three invalid results were observed in the intermediate range because of decreased control line intensity. The minimum sample volume was 30 μL, and maximum hematocrit was 46%. In 40 pregnant and 40 nonpregnant women aged 18-44 years, the device generated 100% concordance with urine qualitative and plasma quantitative test results.CONCLUSIONSThe ADEXUSDx™ hCG test demonstrates acceptable performance for the determination of pregnancy status using capillary fingerstick samples.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom