A Novel System Control for Quality Control of Diagnostic Tests Based on Next-Generation Sequencing
Author(s) -
Maurice Chan,
Arseny Smirnov,
Fabianus Hendriyan Mulawadi,
Priscilla Lim,
Wen-Huey Lim,
See Ting Leong,
Hwee Meng Low,
Mei Qi Yee,
Yong Qiang Yeo,
Xin Zhou,
Charlie Lee,
Wen Huang,
Louis Welebob,
Mengchu Wu
Publication year - 2016
Publication title -
the journal of applied laboratory medicine
Language(s) - English
Resource type - Journals
eISSN - 2576-9456
pISSN - 2475-7241
DOI - 10.1373/jalm.2016.020131
Subject(s) - amplicon , genomic dna , dna sequencing , computational biology , biology , genetics , dna , dna extraction , amplicon sequencing , polymerase chain reaction , gene , 16s ribosomal rna
BACKGROUNDWe describe a novel system control (SC) implemented in an automated AmpliSeq™-based next-generation sequencing (NGS)2 run that simultaneously acts as (a) an external positive/sensitivity control, (b) a spike-in QC for DNA extraction, and (c) a nontemplate control to detect exogenous DNA contamination.METHODSPlasmids carrying wild-type tobacco mosaic virus sequence and a sequence with three designed mutations were synthesized and mixed, such that the mutations are present at 5% variant frequency in the mixture designated as SC. SC was used as a stand-alone sample and spiked into each sample in each run. A cell line-derived reference material, in both a formalin-fixed paraffin-embedded (FFPE) sample and genomic DNA (gDNA), was sequenced in the same runs.RESULTSBy interpolation, 100 fg SC spiked in FFPE sample produced sequencing coverage equivalent to approximately 3 fg in the gDNA. In the SC-only sample, all three designed mutations were recovered around 5% as expected, while no significant reads of human genome were present. In samples with a common PCR inhibitor, coverage for both SC and target amplicons were eliminated. An inverse relationship between the coverage of SC and DNA input was observed. In clinical samples, the ratio of SC to the median coverage of sample can be used to indicate insufficient DNA input.CONCLUSIONSThe SC is an elegant and comprehensive QC concept for NGS-based diagnostic tests.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom