Automated Multiplex LC-MS/MS Assay for Quantifying Serum Apolipoproteins A-I, B, C-I, C-II, C-III, and E with Qualitative Apolipoprotein E Phenotyping
Author(s) -
Irene van den Broek,
Fred P.H.T.M. Romijn,
Jan Nouta,
Arnoud van der Laarse,
Jan W. Drijfhout,
Nico P.M. Smit,
Yuri E. M. van der Burgt,
Christa M. Cobbaert
Publication year - 2015
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1373/clinchem.2015.246702
Subject(s) - apolipoprotein b , multiplex , chemistry , lipoprotein , chromatography , hypertriglyceridemia , microbiology and biotechnology , triglyceride , cholesterol , biology , biochemistry , bioinformatics
Direct and calculated measures of lipoprotein fractions for cardiovascular risk assessment suffer from analytical inaccuracy in certain dyslipidemic and pathological states, most commonly hypertriglyceridemia. LC-MS/MS has proven suitable for multiplexed quantification and phenotyping of apolipoproteins. We developed and provisionally validated an automated assay for quantification of apolipoprotein (apo) A-I, B, C-I, C-II, C-III, and E and simultaneous qualitative assessment of apoE phenotypes.