Mass Spectrometry–Based Escherichia coli H Antigen/Flagella Typing: Validation and Comparison with Traditional Serotyping
Author(s) -
Keding Cheng,
YiMin She,
Huixia Chui,
Larissa Domish,
Angela Sloan,
Drexler Hernandez,
Stuart McCorrister,
Jun Ma,
Bianli Xu,
Aleisha Reimer,
J. David Knox,
Gehua Wang
Publication year - 2016
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1373/clinchem.2015.244236
Subject(s) - typing , serotype , biology , escherichia coli , microbiology and biotechnology , antigen , repeatability , antiserum , genetics , chromatography , chemistry , gene
BACKGROUND Escherichia coli H antigen typing with antisera, a useful method for flagella clinical identification and classification, is a time-consuming process because of the need to induce flagella growth and the occurrence of undetermined strains. We developed an alternative rapid and analytically sensitive mass spectrometry (MS) method, termed MS-based H antigen typing (MS-H), and applied it at the protein sequence level for H antigen typing. We also performed a comparison with traditional serotyping on reference strains and clinical isolates. METHODS On the basis of international guidelines, the analytical selectivity and sensitivity, imprecision, correlation, repeatability, and reproducibility of the MS-H platform was evaluated using reference strains. Comparison of MS-H typing and serotyping was performed using 302 clinical isolates from 5 Canadian provinces, and discrepant results between the 2 platforms were resolved through whole genome sequencing. RESULTS Repeated tests on reference strain EDL933 demonstrated a lower limit of the measuring interval at the subsingle colony (16.97 μg or 1.465 × 107 cells) level and close correlation (r2 > 0.99) between cell culture biomass and sequence coverage. The CV was <10.0% among multiple repeats with 4 reference strains. Intra- and interlaboratory tests demonstrated that the MS-H method was robust and reproducible under various sample preparation and instrumentation conditions. Using discrepancy analysis via whole genome sequencing, performed on isolates with discrepant results, MS-H accurately identified 12.3% more isolates than conventional serotyping. CONCLUSIONS MS-H typing of E. coli is useful for fast and accurate flagella typing and could be very useful during E. coli outbreaks.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom