State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography–Tandem Mass Spectrometry Methods
Author(s) -
Christopher-John L Farrell,
Steven Martin,
Brett McWhinney,
Isabella R. Straub,
Paul F. Williams,
Markus Herrmann
Publication year - 2012
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1373/clinchem.2011.172155
Subject(s) - chromatography , liquid chromatography–mass spectrometry , chemistry , mass spectrometry , tandem mass spectrometry
BACKGROUND Vitamin D testing is increasing worldwide. Recently several diagnostic manufacturers including Abbott and Siemens have launched automated 25-hydroxy vitamin D (25OH-D) immunoassays. Furthermore, preexisting assays from DiaSorin and Roche have recently been modified. We compared the performance of 5 automated immunoassays, an RIA and 2 liquid chromatography–tandem mass spectrometry (LC-MS/MS) methods. METHODS Aliquots of 170 randomly selected patient samples were prepared and 25OH-D was measured by 2 LC-MS/MS methods, an RIA (DiaSorin), and automated immunoassays from Abbott (Architect), DiaSorin (LIAISON), IDS (ISYS), Roche (E170, monoclonal 25OH-D3 assay), and Siemens (Centaur). Within-run and between-run imprecision were evaluated by measurement of 5 replicates of 2 serum pools on 5 consecutive days. RESULTS The LC-MS/MS methods agreed, with a concordance correlation coefficient (CCC) of 0.99 and bias of 0.56 μg/L (1.4 nmol/L). The RIA assay showed a performance comparable to LC-MS/MS, with a CCC of 0.97 and a mean bias of 1.1 μg/L (2.7 nmo/L). All immunoassays measured total 25OH-D (including D3 and D2), with the exception of the Roche assay (D3 only). Among the immunoassays detecting total 25OH-D, the CCCs varied between 0.85 (Abbott) to 0.95 (LIAISON). The mean bias ranged between 0.2 μg/L (0.5 nmol/L) (LIAISON) and 4.56 μg/L (11.4 nmol/L) (Abbott). The Roche 25OH-D3 assay demonstrated small mean bias [−2.7 μg/L (−6.7 nmol/L)] [−2.7 μg/L (−6.7 nmol/L)] but a low CCC of just 0.66. Most assays demonstrated good intra- and interassay precision, with CV <10%. CONCLUSIONS Automated immunoassays demonstrated variable performance and not all tests met our minimum performance goals. It is important that laboratories be aware of the limitations of their assay.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom