Quantification of Urinary Oxalate by Liquid Chromatography–Tandem Mass Spectrometry with Online Weak Anion Exchange Chromatography
Author(s) -
Brian Keevil,
Susan Thornton
Publication year - 2006
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1373/clinchem.2006.075275
Subject(s) - chemistry , chromatography , oxalate , oxalic acid , elution , formic acid , detection limit , mass spectrometry , liquid chromatography–mass spectrometry , tandem mass spectrometry , ion chromatography , ascorbic acid , ammonium oxalate , inorganic chemistry , food science
Background: Urinary oxalate is commonly measured with an enzymatic assay that is specific but requires a manual clean-up step to reduce ascorbic acid interference. We developed a urinary oxalate assay that uses liquid chromatography–tandem mass spectrometry (LC-MS/MS) with anion exchange chromatography and simple sample preparation. Methods: We added calibrator or urine sample (10 μL) to 10 μL of 13C2 oxalate and 400 μL of water and performed separation on a Waters OASIS WAX column, flow rate 0.6 mL/min, and then elution for 0.3 min with water containing 2 mmol/L ammonium acetate and 1 mL/L formic acid and for 1.0 min with 750 mL/L methanol containing 20 mL/L ammonia. We detected multiple reaction monitoring transitions m/z 88.6 > 60.5 and m/z 90.5 > 61.5 for oxalic acid and 13C2-oxalate, respectively, with a Quattro micro tandem mass spectrometer in electrospray-negative mode. Results: Oxalate and 13C2-oxalate eluted at 1.2 min. Mean recovery was 95%, limit of detection 3.0 μmol/L, lower limit of quantification 100.0 μmol/L, linearity to 2212 μmol/L, imprecision <6%, and bias <3% at 166, 880, and 1720 μmol/L. Oxalate eluted after the main area of ion suppression. Mean response ratios for urine and aqueous samples, enriched at 200 and 1000 μmol/L, were 3.7% and 5.4%, respectively. No interference was observed from other organic acids. Passing and Bablock regression analysis comparing the Trinity Biotech enzymatic reagent set and LC-MS/MS showed LC-MS/MS = 1.06 (enzymatic assay) −21.2, r = 0.964, n = 110. Bland Altman analysis showed general agreement, with a mean bias of −1.9 μmol/L. Conclusion: This LC-MS/MS assay is applicable for quantifying urinary oxalate excretion.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom