Rapid and Simultaneous Quantification of Levetiracetam and Its Carboxylic Metabolite in Human Plasma by Liquid Chromatography Tandem Mass Spectrometry
Author(s) -
LiLing Yeap,
YokeLin Lo
Publication year - 2014
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0111544
Subject(s) - chromatography , protein precipitation , chemistry , detection limit , mass spectrometry , formic acid , liquid chromatography–mass spectrometry , tandem mass spectrometry , selected reaction monitoring , analyte , triple quadrupole mass spectrometer , calibration curve
A simple liquid chromatography tandem mass spectrometry method was developed and validated according to the guidelines of the US Food and Drug Administration and the European Medicines Agency for a simultaneous quantification of levetiracetam (LEV) and its metabolite, UCB L057 in the plasma of patients. A 0.050 mL plasma sample was prepared by a simple and direct protein precipitation with 0.450 mL acetonitrile (ACN) containing 1 µg/mL of internal standard (IS, diphenhydramine), then vortex mixed and centrifuged. A 0.100 mL of the clear supernatant was diluted with 0.400 mL water and well mixed. A 0.010 mL of the resultant solution was injected into an Agilent Zorbax SB-C18 (2.1 mm×100 mm, 3.5 µm) column with an isocratic elution at 0.5 mL/min using a mixture of 0.1% formic acid in water and ACN (40∶60 v/v). Detection was performed using an AB Sciex API 3000 triple quadrupole mass spectrometer, equipped with a Turbo Ion Spray source, operating in a positive mode: LEV at transition 171.1>154.1, UCB L057 at 172.5>126.1, and IS at 256.3>167.3; with an assay run time of 2 minutes. The lower limit of quantification (LLOQ) for both LEV and UCB L057 was validated at 0.5 µg/mL, while their lower limit of detection (LOD) was 0.25 µg/mL. The calibration curves were linear between 0.5 and 100 µg/mL for both analytes. The inaccuracy and imprecision of both intra-assay and inter-assay were less than 10%. Matrix effects were consistent between sources of plasma and the recoveries of all compounds were between 100% and 110%. Stability was established under various storage and processing conditions. The carryovers from both LEV and UCB L057 were less than 6% of the LLOQ and 0.13% of the IS. This assay method has been successfully applied to a population pharmacokinetic study of LEV in patients with epilepsy.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom