z-logo
open-access-imgOpen Access
Validation of a Manual Headspace Gas Chromatography Method for Determining Volatile Compounds in Biological Fluids
Author(s) -
Guilherme Vannucchi Portari,
Júlio Sérgio Marchini,
Alceu Afonso Jordão
Publication year - 2007
Publication title -
laboratory medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.332
H-Index - 28
eISSN - 1943-7730
pISSN - 0007-5027
DOI - 10.1309/ea876c1pvc0uxph0
Subject(s) - acetaldehyde , chromatography , gas chromatography , acetone , biological fluids , chemistry , methanol , acetonitrile , ethanol , organic chemistry
Background: We report the validation of a method for the determination of acetaldehyde, acetone, methanol, and ethanol in biological fluids using manual headspace sample introduction and an acetonitrile internal standard. Method: This method uses a capillary column (I = 30 m, I.D. = 0.25 mm, dF = 0.25 mu m) installed in a gas chromatography-flame ionization detector (GC-FID) apparatus with a run time of 7.5 minutes. Results: Analysis of the retention times and the resolution of the analyte peaks demonstrated excellent separation without widening of the peaks. Precision and accuracy were good (interassay precision < 15% and recovery between 85% and 115%) in both blood and urine. Conclusion: The method was linear (r > 0.09) over the analytical measurement range (AMR) of each analyte

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom