Development and Validation of Effective Real-Time and Periodic Interinstrument Comparison Method for Automatic Hematology Analyzers
Author(s) -
Sang Hyuk Park,
ChanJeoung Park,
MiJeong Kim,
Mi-Ok Choi,
Min-Young Han,
YoungUk Cho,
Seongsoo Jang
Publication year - 2014
Publication title -
american journal of clinical pathology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.859
H-Index - 128
eISSN - 1943-7722
pISSN - 0002-9173
DOI - 10.1309/ajcpp0f9qxyoujog
Subject(s) - cutoff , percentile , coefficient of variation , hematology analyzer , statistics , medicine , mathematics , correlation coefficient , nuclear medicine , physics , quantum mechanics
We developed and validated an interinstrument comparison method for automatic hematology analyzers based on the 99th percentile coefficient of variation (CV) cutoff of daily means and validated in both patient samples and quality control (QC) materials.