Premature wear and osteolysis in an HA-coated, uncemented total hip arthroplasty
Author(s) -
Patrick Duffy,
J. L. Sher,
Paul Partington
Publication year - 2004
Publication title -
journal of bone and joint surgery - british volume
Language(s) - English
Resource type - Journals
eISSN - 0968-7300
pISSN - 0301-620X
DOI - 10.1302/0301-620x.86b1.13402
Subject(s) - osteolysis , medicine , prosthesis , asymptomatic , surgery , polyethylene , total hip arthroplasty , radiography , dentistry , retrospective cohort study , materials science , composite material
The ABG I cementless hip prosthesis has demonstrated unacceptably high rates of wear and osteolysis in our patients. We performed a retrospective study of 97 hips implanted between 1992 and 1998. Radiographic analysis revealed high rates of wear of the polyethylene liner with marked periacetabular osteolysis. Clinical examination indicated that many of these patients were initially asymptomatic. Wear-related problems have required ten hips to be revised and a furher 13 are awaiting revision. This gives a failure rate of 24% at a mean follow-up of 69 months. Contributing factors are likely to include poor wear characteristics of the polyethylene liners which were gamma irradiated in air, and increased wear debris caused by a poor fit of the polyethylene liner within the shell. We believe that all ABG I implants should be immediately reviewed and remain under careful, long-term follow-up.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom