Dosimetric Consistency of Co-60 Teletherapy Unit : A Ten Years Study
Author(s) -
Misba Hamid Baba,
M. Mohib-ul-Haq,
Aijaz A. Khan
Publication year - 2013
Publication title -
international journal of health sciences
Language(s) - English
Resource type - Journals
eISSN - 1658-7774
pISSN - 1658-3639
DOI - 10.12816/0006016
Subject(s) - cobalt 60 , dosimetry , nuclear medicine , nuclide , medicine , dose rate , consistency (knowledge bases) , radionuclide , radiation oncologist , medical physics , radiation therapy , radiochemistry , mathematics , physics , irradiation , radiology , chemistry , nuclear physics , geometry
The goal of the Radiation standards and Dosimetry is to ensure that the output of the Teletherapy Unit is within ±2% of the stated one and the output of the treatment dose calculation methods are within ±5%. In the present paper, we studied the dosimetry of Cobalt-60 (Co-60) Teletherapy unit at Sher-I-Kashmir Institute of Medical Sciences (SKIMS) for last 10 years. Radioactivity is the phenomenon of disintegration of unstable nuclides called radionuclides. Among these radionuclides, Cobalt-60, incorporated in Telecobalt Unit, is commonly used in therapeutic treatment of cancer. Cobalt-60 being unstable decays continuously into Ni-60 with half life of 5.27 years thereby resulting in the decrease in its activity, hence dose rate (output). It is, therefore, mandatory to measure the dose rate of the Cobalt-60 source regularly so that the patient receives the same dose every time as prescribed by the radiation oncologist. The under dosage may lead to unsatisfactory treatment of cancer and over dosage may cause radiation hazards. Our study emphasizes the consistency between actual output and output obtained using decay method.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom