Chemiluminescent Enzyme Immunoassay for Measuring Leptin
Author(s) -
Satoshi Sekiguchi,
Hideki Kohno,
Kiyoshi Yasukawa,
Kuniyo Inouye
Publication year - 2011
Publication title -
bioscience biotechnology and biochemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.509
H-Index - 116
eISSN - 1347-6947
pISSN - 0916-8451
DOI - 10.1271/bbb.100885
Subject(s) - leptin , immunoassay , polyclonal antibodies , chemistry , alkaline phosphatase , detection limit , chemiluminescence , bovine serum albumin , antibody , medicine , enzyme , endocrinology , chromatography , biochemistry , immunology , obesity
Leptin is one of the representative adipocyte-derived protein hormones. Measuring the serum leptin concentration gives an important index for preventing and treating diabetes mellitus and other diseases. We constructed in this study a chemiluminescent enzyme immunoassay (CLEIA) for measuring leptin by using the anti-leptin polyclonal antibody and alkaline phosphatase (ALP). The method applies the IgG-conjugated ferrite particle to capture leptin in a sample and the ALP-conjugated Fab fragment to detect the captured leptin. We tested Block ace, CE510, and bovine serum albumin (BSA) for their abilities to block non-specific binding of ALP-conjugated anti-leptin Fab to the ferrite particle and found BSA to be the most effective. The measurable range with this ELISA for leptin was 0.1-1.0 pg/mL of leptin and the detection limit (blank+2SD) was 0.1 pg/mL of leptin. These results demonstrate sufficient sensitivity with our system to measure the serum leptin concentration and its clinical usefulness. The results also suggest that a sensitive enzyme immunoassay can be constructed by using only one polyclonal antibody.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom