The Accuracy of a Near-Infrared Spectroscopy Cerebral Oximetry Device and Its Potential Value for Estimating Jugular Venous Oxygen Saturation
Author(s) -
Keita Ikeda,
David B. MacLeod,
Hilary P. Grocott,
Eugene W. Moretti,
Warwick A. Ames,
Charles Vacchiano
Publication year - 2014
Publication title -
anesthesia and analgesia/anesthesia and analgesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.404
H-Index - 201
eISSN - 1526-7598
pISSN - 0003-2999
DOI - 10.1213/ane.0000000000000463
Subject(s) - medicine , venous blood , anesthesia , pulse oximetry , bland–altman plot , arterial blood , cardiology , nuclear medicine , limits of agreement
An intriguing potential clinical use of cerebral oximeter measurements (SctO2) is the ability to noninvasively estimate jugular bulb venous oxygen saturation (SjvO2). Our purpose in this study was to determine the accuracy of the FORE-SIGHT(®) (CAS Medical Systems, Branford, CT), which is calibrated to a weighted average of 70% (SjvO2) and 30% arterial saturation, for Food and Drug Administration pre-market approval 510(k) certification by adapting an industry standard protocol, ISO 9919:2005 (www.ISO.org) (used for pulse oximeters), and to evaluate the use of SctO2 and SpO2 measurements to noninvasively estimate jugular venous oxygen saturation (SnvO2).