Self-inflating manual resuscitators (SIMRs) can mislead caregivers because the bag, unlike a Mapleson-type device, reinflates even without patient exhalation. We added a whistle as an audible indicator to the exhalation port of a SIMR. In randomized order, each participant provided two sets of breaths via mask ventilation with a SIMR, one with and one without audible feedback, to a Human Patient Simulator modified to log lung volume changes. The last three breaths in each set were used to compare average tidal volume (Vt) under both conditions. Eighty-seven advanced cardiac life support trainees (54 males, 33 females) with clinical experience averaging 6.4 +/- 9.4 yr were recruited. Average Vt delivered with the standard SIMR was 486 +/- 166 mL and 624 +/- 96 mL with the modified SIMR. Average Vt delivered by a modified SIMR was significantly larger by 40% when it followed standard SIMR use and 19% when using the modified SIMR first. Use of a SIMR with an audible indicator of exhalation significantly (P < 0.001) increased mask ventilation of a patient simulator, suggesting that mask ventilation of a patient with a SIMR may also be increased by objective, real-time feedback of exhaled Vt.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom