Steroid Profiling by Tandem Mass Spectrometry Improves the Positive Predictive Value of Newborn Screening for Congenital Adrenal Hyperplasia
Author(s) -
Carla Minutti,
Jean M. Lacey,
Mark J Magera,
Si Houn Hahn,
Mark McCann,
Andreas Schulze,
David Cheillan,
C. Dorche,
Donald H. Chace,
James Lymp,
Donald Zimmerman,
Piero Rinaldo,
Dietrich Matern
Publication year - 2004
Publication title -
the journal of clinical endocrinology and metabolism
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.206
H-Index - 353
eISSN - 1945-7197
pISSN - 0021-972X
DOI - 10.1210/jc.2003-032235
Subject(s) - congenital adrenal hyperplasia , dried blood , immunoassay , predictive value , newborn screening , medicine , hydroxyprogesterone , liquid chromatography–mass spectrometry , spots , endocrinology , chromatography , mass spectrometry , chemistry , steroid , immunology , pediatrics , pathology , antibody , hormone
Congenital adrenal hyperplasia (CAH) is primarily caused by 21-hydroxylase deficiency and leads to an accumulation of 17-hydroxyprogesterone and reduced cortisol levels. Newborn screening for CAH is traditionally based on measuring 17-hydroxyprogesterone by different immunoassays. Despite attempts to adjust cutoff levels for birth weight, gestational age, and stress factors, the positive predictive value for CAH screening remains less than 1%. To improve this situation, we developed a method using liquid chromatography-tandem mass spectrometry to measure 17-hydroxyprogesterone, androstenedione, and cortisol simultaneously in blood spots. A total of 1222 leftover blood spots from six different screening programs using different immunoassays (fluorescent immunoassay and ELISA) were reanalyzed in a blinded fashion by liquid chromatography-tandem mass spectrometry. Thirty-one samples were from babies with CAH, 190 had yielded false-positive results by immunoassay, and the remaining 1001 samples were from babies with normal screening results. Steroid profiling allowed for an elimination of 169 (89%) of the false-positive results and for an improvement of the positive predictive value from the reported 0.5 to 4.7%. Although this method is not suitable for mass screening due to the length of the analysis (12 min), it can be used as a second-tier test of blood spots with positive results for CAH by the conventional methods. This would prevent unnecessary blood draws, medical evaluations, and stress to families.