Background An Ankle Foot Orthosis (AFO) is commonly used in clinical practice to assist gait of patients with different pathologies. The flexibility of the AFO depends on different design characteristics while specific mechanical requirements of the AFO are correlated with patient anatomy and pathology. To this day, the correlation between AFO-design and patient pathology is mainly based on the orthopaedic technician’s experience. The aim of this study is to investigate the influence of the stiffness of an AFO on the muscle-activation pattern of a subject suffering from an anterior muscle insufficiency of the lower leg using a personalized musculoskeletal model. Materials and methods Test subject was a 40-year old male suffering from a hemiparetic anterior muscle insufficiency of the lower leg. A musculoskeletal model of the lower limbs with 23 degrees of freedom and 92 muscles was scaled in OpenSim to match the test subject’s anthropometric data [1]. Muscle-definitions were adapted to simulate the patients’ pathology. A subject specific AFO was constructed using the selective laser sintering technique [2]. The actual stiffness of the AFO was determined using finite element analysis [3] and was 258 Nm/rad. Marker trajectories of an AFO-gait were used to calculate kinematic parameters and muscle-activation during gait using the musculoskeletal model. The AFO was simulated as an angle-dependent torque around the ankle with a neutral angle of 0°. The stiffness of the AFO was varied between 150 and 350 Nm/rad in steps of 25 Nm/rad. Results