A Clinical Evaluation of the Hemocue Haemoglobinometer using Capillary, Venous and Arterial Samples
Author(s) -
P. P. Chen,
Timothy G. Short,
Denis H. Y. Leung,
T. E. Oh
Publication year - 1992
Publication title -
anaesthesia and intensive care
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.494
H-Index - 62
eISSN - 1448-0271
pISSN - 0310-057X
DOI - 10.1177/0310057x9202000419
Subject(s) - medicine , coefficient of variation , venous blood , capillary action , arterial blood , nuclear medicine , anesthesia , chromatography , chemistry , materials science , composite material
The 'Hemocue' device for rapid estimation of haemoglobin concentration was evaluated in a clinical setting. Repeatable accuracy of capillary, venous and arterial samples was examined and then compared with standard laboratory venous haemoglobin estimates using a 'Coulter JT' analyser in 42 patients. The mean values for haemoglobin (g/l) and coefficient of variation were capillary 108.2 (8.0); venous 104.9 (2.2); arterial 105.9 (2.0); and laboratory venous 104.6 (1.3). Although the mean haemoglobin values were similar, capillary samples were significantly less repeatable than venous or arterial samples (Pitman test, P < 0.001). Comparison of variance between the laboratory sample and each sampling technique demonstrated that capillary samples were significantly more variable than venous or arterial samples. Peripheral skin temperature did not influence the accuracy of capillary samples. Hemocue estimations of venous samples were found to be as accurate as laboratory estimations. The lack of repeatable accuracy of capillary estimations was sufficiently large that their use cannot be recommended in clinical practice.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom