Considerations and Recommendations for the Introduction of Objective Performance Criteria for Transcatheter Aortic Heart Valve Device Approval
Author(s) -
Stuart J. Head,
Darren Mylotte,
Michael J. Mack,
Nicolò Piazza,
Nicolas M. Van Mieghem,
Martin B. Leon,
A. Pieter Kappetein,
David R. Holmes
Publication year - 2016
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/circulationaha.115.020493
Subject(s) - medicine , food and drug administration , investigational device exemption , clinical trial , randomized controlled trial , heart valve , risk analysis (engineering) , cardiology , surgery
In the United States, new surgical heart valves can be approved on the basis of objective performance criteria (OPC). In contrast, the US Food and Drug Administration traditionally requires stricter criteria for transcatheter heart valve (THV) approval, including randomized, clinical trials. Recent US Food and Drug Administration approval of new-generation THVs based on single-arm studies has generated interest in alternative study approaches for THV device approval. This review evaluates whether THV device approval could follow a pathway analogous to that of surgical heart valves by incorporating OPC and provides several considerations and recommendations. Factors to be taken into account in the construction of OPC include the maturity of THV technology, variability in transcatheter aortic valve replacement practice, end points included as OPC, follow-up terms for specific OPC, patient populations to which these OPC apply, and (statistical) methods for OPC development. We recommend that approval of THV devices in the United States for low- and intermediate-risk patients or for new indications should provisionally rely on data from randomized, clinical trials. However, it is recommended that formal OPC be applied for approval of new-generation THVs for use in high- and extreme-risk patient populations.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom