Accuracy of a continuous blood pressure monitor based on arterial tonometry.
Author(s) -
Takayuki Sato,
Masanori Nishinaga,
Atsuhiko Kawamoto,
Takeaki Ozawa,
Hiroyoshi Takatsuji
Publication year - 1993
Publication title -
hypertension
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.986
H-Index - 265
eISSN - 1524-4563
pISSN - 0194-911X
DOI - 10.1161/01.hyp.21.6.866
Subject(s) - supine position , blood pressure , medicine , valsalva maneuver , cardiology , continuous noninvasive arterial pressure , diastole , waveform , anesthesia , physics , quantum mechanics , voltage
A validation study of the continuous noninvasive tonometric blood pressure monitor called JENTOW was performed in 20 normotensive subjects and 10 hypertensive patients. Tonometric and intra-arterial blood pressures were simultaneously recorded at supine rest and during a Valsalva maneuver and tilting test. The results of the strict evaluation of the instrument's capacity for reproducing intra-arterial blood pressure were as follows: 1) The overall frequency response of the transcutaneous blood pressure-monitoring system based on arterial tonometry was flat, with negligible delay to intra-arterial blood pressure in the range of 0-5 Hz. 2) The largest discrepancy between intra-arterial and tonometric pressure waveforms was found at the early systolic phase; except for this phase, the tonometric waveform was almost equal to the intra-arterial waveform. 3) The beat-to-beat variability of tonometric pressure corresponded to that of intra-arterial pressure almost perfectly in the physiologically significant frequency range of 0-0.5 Hz. 4) During resting conditions, the averages of the systolic and diastolic values measured tonometrically corresponded well to those measured intra-arterially. 5) The changes in the between-method discrepancy of blood pressure values during the Valsalva maneuver were statistically significant but small (< 5 mm Hg). 6) No significant effect of postural tilting was found on the between-method discrepancy. We conclude that this method is clinically acceptable and reliable except for its limited capacity for recording the higher frequency intra-arterial waveform and for responding to the relatively rapid and large transient changes in blood pressure.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom