Internal cardiac defibrillation in man: pronounced improvement with sequential pulse delivery to two different lead orientations.
Author(s) -
Douglas L. Jones,
George J. Klein,
G Guiraudon,
Arjun D. Sharma,
Michael J. Kallok,
J. D. Bourland,
Willis A. Tacker
Publication year - 1986
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/01.cir.73.3.484
Subject(s) - defibrillation , medicine , cardiology , shock (circulatory) , defibrillation threshold , catheter , ventricle , pulse (music) , cardiac output , surgery , hemodynamics , voltage , electrical engineering , engineering
Wider applicability of an implantable automatic defibrillator depends on achieving internal cardiac defibrillation consistently with the lowest possible energy. In animal studies, we have found that the cardiac defibrillation threshold could be reduced when sequential shocks separated in time and spacially arranged were delivered to the heart. We compared internal cardiac defibrillation using a single pulse shock delivered through an intravascular catheter with this new method for internal cardiac defibrillation in patients undergoing cardiac surgery for the correction of arrhythmias. For the single pulse shock and the first pulse of the sequential pulse shock, current was passed through an intravascular catheter with the catheter cathode at the apex of the right ventricle and the anode at the superior vena cava-atrial junction region. The second pulse of the sequential pulse countershock was delivered between the catheter cathode in the right ventricular apex and an oval plaque electrode secured on the laterobasal left ventricular epicardium as anode. With the single pulse alone for shock delivery, 12 patients could be defibrillated with an average of 20.1 +/- 16.8 J, with a corresponding leading-edge peak voltage and current of 836 +/- 319 V and 9.4 +/- 4.5 A, respectively. However, two of the patients could not be defibrillated with energies below 50 J. With the sequential pulse shock delivery, a significant reduction in all values were recorded. Mean total energy for defibrillation averaged 7.7 +/- 6.0 J. Leading-edge peak voltage and current from the catheter averaged 430 +/- 148 V and 5.0 +/- 2.8 A, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom