Inhibition of bipolar demand pacemaker by diaphragmatic myopotentials.
Author(s) -
S. Serge Barold,
Ling S. Ong,
Michael D. Falkoff,
Robert A. Heinle
Publication year - 1977
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/01.cir.56.4.679
Subject(s) - medicine , diaphragmatic breathing , cardiology , anesthesia , pathology , alternative medicine
This report describes inhibition of a normally functioning bipolar demand pulse generator by diaphragmatic myopotentials. Transient pacemaker suppression occurred repeatedly with deep respiration, straining, the Valsalva maneuver, coughing, sneezing and laughing. When the magnet was applied, none of these maneuvers inhibited the pacemaker. Extensive investigations ruled out an intermittent electrode problem such as a wire fracture or insulation break. Sensing of diaphragmatic myopotentials should be considered in the differential diagnosis of unexplained pacemaker pauses.