Fluoroscope-generated electromagnetic interference in an external demand pacemaker: report of a case.
Author(s) -
C J Leeds,
Masood Akhtar,
A N Damato
Publication year - 1977
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/01.cir.55.3.548
Subject(s) - medicine , fluoroscopy , electromagnetic interference , interference (communication) , pulse generator , on demand , cardiac pacemaker , cardiology , electrical engineering , surgery , voltage , computer science , engineering , channel (broadcasting) , multimedia
Electromagnetic interference presented as inhibition and resetting of the demand circuitry of a ventricular-inhibited temporary external pacemaker in a 70-year-old man undergoing surgical implantation of a permanent bipolar pacemaker generator and lead. The arrhythmia was found to be due to oversensing of waveforms modulated to simulate myocardial potentials emitted by a faulty fluoroscopy unit used in the vicinity of external temporary transvenous pacing equipment. The documentation of this disruption of pacemaker rhythm reinforces the need for continuous monitoring of patients treated with external demand pacemakers and for the careful maintenance of all electrical equipment.