z-logo
open-access-imgOpen Access
Impaired Function of Cloth-Covered Starr-Edwards Mitral Valve Prosthesis
Author(s) -
Ernest Craige,
Peter Hutchin,
Richard Sutton
Publication year - 1970
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/01.cir.41.1.141
Subject(s) - medicine , cardiology , ball valve , prosthesis , mitral valve , thrombus , surgery , structural engineering , engineering
Malfunction of a cloth-covered Starr-Edwards mitral valve prosthesis (6300 series) was suspected because of phonocardiographic abnormalities. A prolonged interval from aortic closure to opening of the mitral ball suggested an impediment to movement of the ball. This was confirmed by catheterization and cinefluorography. At autopsy the impaired ball movement was found to be due to thrombus buildup on the outflow face of the mitral valve orifice. It is suggested that frequent postoperative phonocardiographic observations may be useful in detecting malfunction of valvular prostheses.