Comparative Study of a Portable Prothrombin Time Monitor Employing Three Different Systems in Oral Anticoagulant Units
Author(s) -
M. Vacas,
María Fernández,
F Martínez-Brotóns,
P.J. Lafuente,
Francisco Ripoll,
Carolina Rodríguez Álvarez,
J.A. Iriarte
Publication year - 2001
Publication title -
pathophysiology of haemostasis and thrombosis
Language(s) - English
Resource type - Journals
eISSN - 1424-8840
pISSN - 1424-8832
DOI - 10.1159/000048040
Subject(s) - prothrombin time , oral anticoagulant , medicine , anticoagulant , warfarin , whole blood , nuclear medicine , anesthesia , surgery , atrial fibrillation
The aim of this study was to evaluate the accuracy of the portable coagulometer CoaguChek (Roche Diagnostics) as a prothrombin time (PT) monitor, and to correlate capillary blood results with those of three different routine methods used for monitoring oral anticoagulant therapy (OAT): capillary, plasma and whole blood samples. Three hospitals participated in the study with a total of 235 patients on OAT. The international normalized ratio (INR) results obtained with CoaguChek were compared with those obtained using each of the routine methods. The study presents a good correlation between the PT monitor and the three methods studied: r = 0.9745 (hospital A), r = 0.9283 (hospital B), r = 0.9136 (hospital C). A simplified concordance test of the methods results in a nine-field comparison table showing concordances of 87.2, 85.7 and 68.4%, respectively. The absolute difference (mean +/- SD) between laboratory A and CoaguChek INRs was 0.0571 +/- 0.2042, with values of 0.04286 +/- 0.3906 for laboratory B and 0.6986 +/- 0.6170 for laboratory C. These results confirm that CoaguChek could be used as a new method for oral anticoagulant monitoring, and is in best agreement with the capillary blood PT system.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom