Comparing School-Aged Refraction Measurements Using the 2WIN-S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross-Sectional Study
Author(s) -
Ziming Liu,
Emmanuel Eric Pazo,
Hong Ye,
Cui Yu,
Ling Xu,
Wei He
Publication year - 2021
Publication title -
journal of ophthalmology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.818
H-Index - 40
eISSN - 2090-0058
pISSN - 2090-004X
DOI - 10.1155/2021/6612476
Subject(s) - retinoscopy , repeatability , cycloplegia , medicine , limits of agreement , intraclass correlation , subjective refraction , standard deviation , confidence interval , mean difference , refraction , optometry , ophthalmology , physics , optics , mathematics , refractive error , statistics , nuclear medicine , eye disease , clinical psychology , psychometrics
Purpose To assess the repeatability and agreement of refractive measurements using 2WIN-S photoscreening with the gold-standard cycloplegic retinoscope refraction.Design Single centre, cross-sectional study.Methods Spherical, cylindrical, axis, and spherical equivalent of 194 bilateral eyes of 97 children were assessed using a retinoscope and 2WIN-S. One week later, another operator repeated the 2WIN-S measurements. The primary outcome measures were to assess the repeatability and agreement between spherical equivalent, J0, and J45 readings of 2WIN-S. The repeatability of measurements was assessed by the within-subject standard deviation (2.77 Sw) and intraclass correlation coefficient (ICC). The agreement between devices was assessed using 95% limits of agreement. The extent of the agreement between cycloplegic retinoscopy and noncycloplegic 2WIN-S measurements was assessed using Bland–Altman analysis.Results The mean age ± SD was 10.3 ± 2.46 year (range, 4–14 years). The sphere, cylinder, and spherical equivalent measurements were found to be consistent with both apparatus ( r value >0.86). ICC for SE, J0, and J45 was 0.900, 0.666, and 0.639, respectively; Sw for SE, J0, and J45 was 0.61D, 0.30D, and 0.31D, respectively; Bland–Altman analysis of retinoscopy with cycloplegia and 2WIN-S for SE was 184/194 (95%) in 95% confidence interval, and the mean value was 0.46. J0 was 184/194 (95%), and the mean value is −0.04. J45 was 181/194 (93%), and the mean value is −0.15.Conclusion The objective refractive measurement of 2WIN-S had good reliability and high agreement with the gold-standard retinoscopy refraction in children and adolescents. While consistency was observed, it is essential to take into consideration that it is a screening tool.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom