Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
Author(s) -
C LÓPEZ DE LA FUENTE,
Ana SánchezCano,
Francisco Segura,
Lorena FuentesBroto,
Isabel Pinilla
Publication year - 2014
Publication title -
biomed research international
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.772
H-Index - 126
eISSN - 2314-6141
pISSN - 2314-6133
DOI - 10.1155/2014/808646
Subject(s) - repeatability , scheimpflug principle , intraclass correlation , anterior chamber angle , reproducibility , ophthalmology , coefficient of variation , correlation coefficient , medicine , corneal pachymetry , mathematics , interclass correlation , biomedical engineering , cornea , materials science , intraocular pressure , corneal topography , statistics
Purpose . To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods . Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results . Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions . GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom