New Estimators and Guidelines for Better Use of Fetal Heart Rate Estimators with Doppler Ultrasound Devices
Author(s) -
Iulian Voicu,
Sébastien Ménigot,
Denis Kouamé,
JeanMarc Girault
Publication year - 2014
Publication title -
computational and mathematical methods in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.462
H-Index - 48
eISSN - 1748-6718
pISSN - 1748-670X
DOI - 10.1155/2014/784862
Subject(s) - estimator , doppler effect , doppler ultrasound , autocorrelation , fetal heart rate , sensitivity (control systems) , ultrasound , statistics , mathematics , medicine , computer science , heart rate , radiology , physics , engineering , electronic engineering , astronomy , blood pressure
Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.