Hemodynamic Surveillance of Ventricular Pacing Effectiveness with the Transvalvular Impedance Sensor
Author(s) -
Valéria Calvi,
G. Pizzimenti,
Marco Lisi,
G Doria,
Ludovico Vasquez,
Francesco Lisi,
S Felis,
Donatella Tempio,
Alfredo Virgilio,
A. Barbetta,
Franco Di Gregorio
Publication year - 2014
Publication title -
advances in medicine
Language(s) - English
Resource type - Journals
eISSN - 2356-6752
pISSN - 2314-758X
DOI - 10.1155/2014/307168
Subject(s) - medicine , cardiology , supine position , hemodynamics
The Transvalvular Impedance (TVI) is derived between atrial and ventricular pacing electrodes. A sharp TVI increase in systole is an ejection marker, allowing the hemodynamic surveillance of ventricular stimulation effectiveness in pacemaker patients. At routine follow-up checks, the ventricular threshold test was managed by the stimulator with the supervision of a physician, who monitored the surface ECG. When the energy scan resulted in capture loss, the TVI system must detect the failure and increase the output voltage. A TVI signal suitable to this purpose was present in 85% of the tested patients. A total of 230 capture failures, induced in 115 patients in both supine and sitting upright positions, were all promptly recognized by real-time TVI analysis (100% sensitivity). The procedure was never interrupted by the physician, as the automatic energy regulation ensured full patient's safety. The pulse energy was then set at 4 times the threshold to test the alarm specificity during daily activity (sitting, standing up, and walking). The median prevalence of false alarms was 0.336%. The study shows that TVI-based ejection assessment is a valuable approach to the verification of pacing reliability and the autoregulation of ventricular stimulation energy.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom