Comparison of Optical versus Ultrasonic Biometry in Keratoconic Eyes
Author(s) -
Yasin Çınar,
Abdullah Kürşat Cingü,
Muhammed Şahin,
Alparslan Şahin,
Harun Yüksel,
Fatih Mehmet Türkçü,
Tuba Çınar,
İhsan Çaça
Publication year - 2013
Publication title -
journal of ophthalmology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.818
H-Index - 40
eISSN - 2090-0058
pISSN - 2090-004X
DOI - 10.1155/2013/481238
Subject(s) - medicine , ultrasonic sensor , ophthalmology , optometry , radiology
Purpose . To compare the measurements of optical versus ultrasonic biometry devices in keratoconic eyes. Materials and Methods . Forty-two eyes of 42 keratoconus (KC) patients enrolled in the study were examined. Clinical and demographic characteristics of the patients were noted, and detailed ophthalmological examination was performed. Following Pentacam measurements, central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were obtained using the Lenstar and US biometer to determine the reproducibility of the measurements between the two devices in keratoconic eyes. The Bland-Altman method was used to describe the agreement between the two devices. Results . The Lenstar could not measure at least one of the biometric properties in one eye and did not automatically give the corrected ACD in 2/3 of our study population. The Lenstar measured CCT (average difference 5.4 ± 19.6 µ m; ICC = 0.90; P < 0.001), LT (average difference 0.13 ± 0.17 mm; ICC = 0.67; P < 0.001), and AL (average difference 0.10 ± 0.76 mm; ICC = 0.75; P < 0.001) thinner than US biometer, whereas it measured ACD (average difference 0.18 ± 0.17 mm; ICC = 0.85; P < 0.001) deeper than US biometer in keratoconic eyes. Conclusion . Although the difference between the measurements obtained using the two devices might be clinically acceptable, US biometry and Lenstar should not be used interchangeably for biometric measurements in KC patients.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom