A Novel CT Imaging System with Adjacent Double X-Ray Sources
Author(s) -
Mou An,
Yaoqin Xie
Publication year - 2013
Publication title -
computational and mathematical methods in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.462
H-Index - 48
eISSN - 1748-6718
pISSN - 1748-670X
DOI - 10.1155/2013/391212
Subject(s) - x ray tube , projection (relational algebra) , computer science , artifact (error) , tomography , computer vision , x ray , computed tomography , power (physics) , artificial intelligence , computer graphics (images) , optics , physics , medicine , radiology , algorithm , electrode , quantum mechanics , anode
Current computed tomography (CT) scanners rotate fast to reduce motion artifact. X-ray tube must work in a high power to make the image clear under short exposure time. However, the life span of such a tube may be shortened. In this paper, we propose a novel double sources CT imaging system, which puts two of the same X-ray sources closely with each other. The system is different from current dual source CT with orthogonal X-ray sources. In our system, each projection is taken twice by these two sources to enhance the exposure value and then recovered to a single source projection for image reconstruction. The proposed system can work like normal single source CT system, while halving down the working power for each tube.