Thrombocytopenia in the Setting of Hemodialysis Using Biocompatible Membranes
Author(s) -
Kathryn B. Muir,
Clifford D. Packer
Publication year - 2012
Publication title -
case reports in medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.2
H-Index - 20
eISSN - 1687-9627
pISSN - 1687-9635
DOI - 10.1155/2012/358024
Subject(s) - biocompatible material , polysulfone , hemodialysis , membrane , dialysis , medicine , polyethylene glycol , biomedical engineering , chemistry , organic chemistry , biochemistry
Thrombocytopenia is a known potential side effect of hemodialysis, however, it is rarely seen in patients who undergo hemodialysis using biocompatible membranes. This case demonstrates hemodialysis-associated thrombocytopenia with use of biocompatible dialysis membranes that expose blood directly to polysulfone. The thrombocytopenia resolved in this patient when the dialysis membrane was changed to a biocompatible model with a polyethylene glycol barrier layer preventing direct interaction between patient blood and polysulfone. The calculated Naranjo ADR score of 9 indicates a highly probable adverse reaction.