Right ventricular dysfunction in chronic thromboembolic obstruction of the pulmonary artery: a pressure-volume study using the conductance catheter
Author(s) -
Colm McCabe,
Paul A. White,
Stephen P. Hoole,
Richard G. Axell,
Andrew N. Priest,
Deepa Gopalan,
Dolores Taboada,
Robert Ross,
Nicholas W. Morrell,
Leonard M. Shapiro,
Joanna PepkeŻaba
Publication year - 2013
Publication title -
journal of applied physiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.253
H-Index - 229
eISSN - 8750-7587
pISSN - 1522-1601
DOI - 10.1152/japplphysiol.01123.2013
Subject(s) - cardiology , medicine , stroke volume , afterload , pulmonary artery , pulmonary hypertension , vascular resistance , blood pressure , ejection fraction , heart failure
Pressure-volume loops describe dynamic ventricular performance, relevant to patients with and at risk of pulmonary hypertension. We used conductance catheter-derived pressure-volume loops to measure right ventricular (RV) mechanics in patients with chronic thromboembolic pulmonary arterial obstruction at different stages of pathological adaptation. Resting conductance catheterization was performed in 24 patients: 10 with chronic thromboembolic pulmonary hypertension (CTEPH), 7 with chronic thromboembolic disease without pulmonary hypertension (CTED), and 7 controls. To assess the validity of conductance measurements, RV volumes were compared in a subset of 8 patients with contemporaneous cardiac magnetic resonance (CMR). Control, CTED, and CTEPH groups showed different pressure-volume loop morphology, most notable during systolic ejection. Prolonged diastolic relaxation was seen in patients with CTED and CTEPH [tau = 56.2 ± 6.7 (controls) vs. 69.7 ± 10.0 (CTED) vs. 67.9 ± 6.2 ms (CTEPH), P = 0.02]. Control and CTED groups had lower afterload (Ea) and contractility (Ees) compared with the CTEPH group (Ea = 0.30 ± 0.10 vs. 0.52 ± 0.24 vs. 1.92 ± 0.70 mmHg/ml, respectively, P < 0.001) (Ees = 0.44 ± 0.20 vs. 0.59 ± 0.15 vs. 1.13 ± 0.43 mmHg/ml, P < 0.01) with more efficient ventriculoarterial coupling (Ees/Ea = 1.46 ± 0.30 vs. 1.27 ± 0.36 vs. 0.60 ± 0.18, respectively, P < 0.001). Stroke volume assessed by CMR and conductance showed closest agreement (mean bias +9 ml, 95% CI -1 to +19 ml) compared with end-diastolic volume (+48 ml, -16 to 111 ml) and end-systolic volume (+37 ml, -21 to 94 ml). RV conductance catheterization detects novel alteration in pressure-volume loop morphology and delayed RV relaxation in CTED, which distinguish this group from controls. The observed agreement in stroke volume assessed by CMR and conductance suggests RV mechanics are usefully measured by conductance catheter in chronic thromboembolic obstruction.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom