A LED-based phosphorimeter for measurement of microcirculatory oxygen pressure
Author(s) -
Philippe Guerci,
Yasin Ince,
Paul M. Heeman,
Dirk J. Faber,
Bülent Ergin,
Can İnce
Publication year - 2016
Publication title -
journal of applied physiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.253
H-Index - 229
eISSN - 8750-7587
pISSN - 1522-1601
DOI - 10.1152/japplphysiol.00316.2016
Subject(s) - phosphorescence , microsecond , quenching (fluorescence) , chemistry , flash (photography) , oxygen , diode , fluorescence , oxygenation , materials science , analytical chemistry (journal) , biomedical engineering , optoelectronics , optics , chromatography , anesthesia , physics , medicine , organic chemistry
Quantitative measurements of microcirculatory and tissue oxygenation are of prime importance in experimental research. The noninvasive phosphorescence quenching method has given further insight into the fundamental mechanisms of oxygen transport to healthy tissues and in models of disease. Phosphorimeters are devices dedicated to the study of phosphorescence quenching. The experimental applications of phosphorimeters range from measuring a specific oxygen partial pressure (Po 2 ) in cellular organelles such as mitochondria, finding values of Po 2 distributed over an organ or capillaries, to measuring microcirculatory Po 2 changes simultaneously in several organ systems. Most of the current phosphorimeters use flash lamps as a light excitation source. However, a major drawback of flash lamps is their inherent plasma glow that persists for tens of microseconds after the primary discharge. This complex distributed excitation pattern generated by the flash lamp can lead to inaccurate Po 2 readings unless a deconvolution analysis is performed. Using light-emitting diode (LED), a rectangular shaped light pulse can be generated that provides a more uniformly distributed excitation signal. This study presents the design and calibration process of an LED-based phosphorimeter (LED-P). The in vitro calibration of the LED-P using palladium(II)-meso-tetra(4-carboxyphenyl)-porphyrin (Pd-TCCP) as a phosphorescent dye is presented. The pH and temperature were altered to assess whether the decay times of the Pd-TCCP measured by the LED-P were significantly influenced. An in vivo validation experiment was undertaken to measure renal cortical Po 2 in a rat subjected to hypoxic ventilation conditions and ischemia/reperfusion. The benefits of using LEDs as a light excitation source are presented.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom