Comparison of the BD Veritor System for Flu A+B with the Alere BinaxNOW Influenza A&B Card for Detection of Influenza A and B Viruses in Respiratory Specimens from Pediatric Patients
Author(s) -
Ferdaus Hassan,
Ashley Nguyen,
Ashley Formanek,
James J. Bell,
Rangaraj Selvarangan
Publication year - 2014
Publication title -
journal of clinical microbiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.349
H-Index - 255
eISSN - 1070-633X
pISSN - 0095-1137
DOI - 10.1128/jcm.02484-13
Subject(s) - medicine , gold standard (test) , virology , respiratory system , virus , confidence interval , influenza a virus , real time polymerase chain reaction , biology , gene , biochemistry
The performance characteristics of two commercially available rapid tests for influenza, the BD Veritor System for Flu A+B (BD) and the Alere BinaxNOW influenza A&B card (BN), were evaluated using 200 frozen clinical specimens collected from January 2011 to June 2012 from pediatric patients. Real-time reverse transcriptase PCR (RT-PCR) was used as the gold standard to evaluate the results obtained by the two different assays. Of the 200 specimens tested, real-time RT-PCR assay detected influenza A or B virus in 116 samples, while BD detected 104 samples and BN detected 84 samples as positive. The overall sensitivity and specificity for detection of both influenza A and B virus in comparison to those of real-time RT-PCR were 89.6% (95% confidence interval [CI], 82.2 to 94.3) and 98.8% (95% CI, 92.6 to 99.9) for BD Veritor and 72.4% (95% CI, 63.2 to 80.0) and 100% (95% CI, 94.5 to 100.0) for BinaxNOW. Workflow analysis indicated that overall processing times for a batch size of 10 specimens were virtually identical between both systems. Overall, these results indicate that the BD Veritor assay was more sensitive than the BinaxNOW assay in detection of influenza A and B viruses in respiratory specimens from pediatric patients.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom