Comparison of the Abbott RealTi m e CT New Formulation Assay with Two Other Commercial Assays for Detection of Wild-Type and New Variant Strains of Chlamydia trachomatis
Author(s) -
Niels FrimodtMøller,
Lisbeth Nørum Pedersen,
Kenneth Persson
Publication year - 2009
Publication title -
journal of clinical microbiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.349
H-Index - 255
eISSN - 1070-633X
pISSN - 0095-1137
DOI - 10.1128/jcm.01446-09
Subject(s) - chlamydia trachomatis , taqman , plasmid , virology , biology , microbiology and biotechnology , chlamydiaceae , 23s ribosomal rna , chlamydia , real time polymerase chain reaction , gene , genetics , rna , ribosome
In an analytical-method comparison study of clinical samples, the Abbott RealTime CT new formulation assay (m2000 real-time PCR), consisting of a duplex PCR targeting different parts of the cryptic plasmid in Chlamydia trachomatis, was compared both with version 2 of the Roche Cobas TaqMan CT assay, comprising a duplex PCR for a target in the cryptic plasmid and the omp1 gene, and with the Gen-Probe Aptima Combo 2 assay (AC2) targeting the C. trachomatis 23S rRNA molecule. First-catch urine samples from Sweden were tested in Malmö, Sweden, for C. trachomatis with the m2000 real-time PCR assay and with an in-house PCR for the new variant C. trachomatis strain with a deletion in the cryptic plasmid. Aliquots of the urine samples were sent to Aarhus, Denmark, where they were further examined with the TaqMan CT and AC2 assays. A positive prevalence of 9.1% (148/1,632 urine samples examined) was detected according to the combined reference standard. The sensitivities and specificities of the three assays were as follows: for the Abbott m2000 assay, 95.3% (141/148) and 99.9% (1,483/1,485), respectively; for the Roche TaqMan assay, 82.4% (122/148) and 100.0% (1,485/1,485); and for the Gen-Probe AC2 assay, 99.3% (147/148) and 99.9% (1,484/1,485). The plasmid mutant strain was detected in 24% (36/148) of the C. trachomatis-positive samples. There is a difference in sensitivity between the new formulations of the Abbott and the Roche assays, but both assays detected the wild-type and new variant C. trachomatis strains equally well.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom