z-logo
open-access-imgOpen Access
Comparison of the Abbott m 2000 RealTi m e CT Assay and the Cepheid GeneXpert CT/NG Assay to the Roche Amplicor CT Assay for Detection of Chlamydia trachomatis in Ocular Samples from Tanzania
Author(s) -
Laura Dize,
Sheila K. West,
James A. Williams,
Barbara Van Der Pol,
Thomas C. Quinn,
Charlotte A. Gaydos
Publication year - 2013
Publication title -
journal of clinical microbiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.349
H-Index - 255
eISSN - 1070-633X
pISSN - 0095-1137
DOI - 10.1128/jcm.00519-13
Subject(s) - genexpert mtb/rif , chlamydia trachomatis , confidence interval , medicine , virology , chlamydia , chlamydiaceae , chlamydiales , biology , tuberculosis , immunology , pathology , mycobacterium tuberculosis
The GeneXpert CT/NG assay (GeneXpert) and the Abbott m2000 RealTime CT (m2000) assay were compared to Amplicor for detecting ocular Chlamydia trachomatis. Discordant specimens were tested by the Aptima CT assay. The m2000 assay sensitivity was 100% (95% confidence interval [CI], 90% to 100%), and specificity was 98.46% (95% CI, 95.2% to 99.2%); GeneXpert sensitivity was 100% (95% CI, 90% to 100%), and specificity was 100% (95% CI, 98.1% to 100%). The m2000 and GeneXpert assays appear to perform as well as the Amplicor assay.