New Rapid Bioassay of Gentamicin Based on Luciferase Assay of Extracellular ATP in Bacterial Cultures
Author(s) -
Lennart Nilsson
Publication year - 1978
Publication title -
antimicrobial agents and chemotherapy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.07
H-Index - 259
eISSN - 1070-6283
pISSN - 0066-4804
DOI - 10.1128/aac.14.6.812
Subject(s) - gentamicin , bioassay , extracellular , luciferase , aminoglycoside , microbiology and biotechnology , escherichia coli , antibiotics , agar , chemistry , biology , bacteria , biochemistry , pharmacology , transfection , genetics , gene
Bacterial ATP levels inEscherichia coli cultures were determined by luciferin/luciferase assay. Rapid effects of gentamicin on ATP levels in the cultures were studied, and a dose-dependent accumulation of extracellular ATP was demonstrated. This phenomenon was used to develop a rapid bioassay of gentamicin in clinical serum specimens. The accuracy of the assay, expressed as coefficient of variation over the range 16 to 1 μg/ml, varied between 1.8 and 6.5%, compared to 2.3 to 4.7% for an agar disk diffusion assay. Serum gentamicin concentrations were determined by both methods in 122 serum specimens from 43 patients receiving gentamicin alone or in combination with other antibiotics, and the results were compared (r = 0.942). Determination of gentamicin based on luciferase assay of extracellular ATP in bacterial cultures requires only 10 μl of serum, allowing for capillary sampling, and results are available within 2 h.