Design of a Magnetic Resonance Imaging Guided Magnetically Actuated Steerable Catheter
Author(s) -
Taoming Liu,
Nate Lombard Poirot,
Tipakorn Greigarn,
M. Cenk Çavuşoğlu
Publication year - 2017
Publication title -
journal of medical devices
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.242
H-Index - 29
eISSN - 1932-619X
pISSN - 1932-6181
DOI - 10.1115/1.4036095
Subject(s) - ablation , computer science , catheter , magnetic resonance imaging , cardiac ablation , actuator , contact force , simulation , catheter ablation , engineering , physics , medicine , surgery , radiology , aerospace engineering , quantum mechanics , artificial intelligence
This paper presents design optimization of a magnetic resonance imaging (MRI) actuated steerable catheter for atrial fibrillation ablation in the left atrium. The catheter prototype, built over polymer tubing, is embedded with current-carrying electromagnetic coils. The prototype can be deflected to a desired location by controlling the currents passing through the coils. The design objective is to develop a prototype that can successfully accomplish the ablation task. To complete the tasks, the catheter needs to be capable of reaching a set of desired targets selected by a physician on the chamber and keeping a stable contact with the chamber surface. The design process is based on the maximization of the steering performance of the catheter by evaluating its workspace in free space. The selected design is validated by performing a simulation of an ablation intervention on a virtual model of the left atrium with a real atrium geometry. This validation shows that the prototype can reach every target required by the ablation intervention and provide an appropriate contact force against the chamber.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom