Circuit Oxygenator Contributes to Extracorporeal Membrane Oxygenation–Induced Hemolysis
Author(s) -
D. Pauline Williams,
Jennifer L. Turi,
Christoph P. Hornik,
Desiree Bonadonna,
Walter L. Williford,
Richard Walczak,
Kevin M. Watt,
Ira M. Cheifetz
Publication year - 2015
Publication title -
asaio journal
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 0.961
H-Index - 66
eISSN - 1538-943X
pISSN - 1058-2916
DOI - 10.1097/mat.0000000000000173
Subject(s) - oxygenator , hemolysis , extracorporeal membrane oxygenation , medicine , oxygenation , membrane oxygenator , anesthesia , cardiology , cardiopulmonary bypass
Hemolysis can occur as a consequence of extracorporeal membrane oxygenation (ECMO) and is associated with increased mortality and morbidity. Shear stress generated by flow through the circuit and oxygenator is believed to cause ECMO-induced hemolysis. We hypothesize that either a smaller dimension oxygenator or an in-line hemofilter will increase ECMO-associated hemolysis. Circuits were configured with a Quadrox-D Adult oxygenator (surface area 1.8 m), Quadrox-iD Pediatric oxygenator (surface area 0.8 m), or Quadrox-D Adult oxygenator with an in-line hemofilter (N = 4) and ran for 6 hours. Samples were collected hourly from the ECMO circuit and a time-based hemolysis control. Plasma hemoglobin levels were assayed. Circuit-induced hemolysis at each time point was defined as the change in plasma hemoglobin standardized to the time-based hemolysis control. Plasma hemoglobin increased with the use of the smaller dimension pediatric oxygenator as compared with the adult oxygenator when controlling for ECMO run time (p = 0.02). Furthermore, there was a greater pressure gradient with the smaller dimension pediatric oxygenator (p < 0.05). Plasma hemoglobin did not change with the addition of the in-line hemofilter. The use of a smaller dimension pediatric oxygenator resulted in greater hemolysis and a higher pressure gradient. This may indicate that the increased shear forces augment ECMO-induced hemolysis.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here