z-logo
open-access-imgOpen Access
Assessment of haemodialysis adequacy by ionic dialysance: intra-patient variability of delivered treatment
Author(s) -
Christopher W. McIntyre
Publication year - 2003
Publication title -
nephrology dialysis transplantation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.654
H-Index - 168
eISSN - 1460-2385
pISSN - 0931-0509
DOI - 10.1093/ndt/18.3.559
Subject(s) - medicine , hemodialysis , urology , urea , blood sampling , surgery , chemistry , organic chemistry
Adequate delivered dose of solute removal (as assessed by urea reduction and calculation of Kt/V) is an important determinant of clinical outcome in chronic haemodialysis (HD) patients. The requirement for multiple blood sampling and efforts taken to minimize the effects of rebound on post-treatment samples ensure Kt/V is measured only intermittently. On-line conductivity monitoring (using sodium flux as a surrogate for urea) allows the repeated non-invasive measurement of Kt/V on each HD treatment. We have studied the accuracy of this method of measuring Kt/V, and the variability of treatment dose delivered to individual patients.