Detection of JWH-018 and JWH-073 by UPLC-MS-MS in Postmortem Whole Blood Casework
Author(s) -
Kevin G. Shanks,
T. Dahn,
Andrea R. Terrell
Publication year - 2012
Publication title -
journal of analytical toxicology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.161
H-Index - 76
eISSN - 1945-2403
pISSN - 0146-4760
DOI - 10.1093/jat/bks013
Subject(s) - chemistry , chromatography , synthetic cannabinoids , analyte , detection limit , electrospray ionization , mass spectrometry , tandem mass spectrometry , liquid chromatography–mass spectrometry , biochemistry , receptor , cannabinoid
Synthetic cannabinoids have been detected in various herbal blends sold legally in convenience stores, smoke shops, and on the Internet. Many of these compounds have extreme forensic significance. We developed and validated a rapid ultra-performance liquid chromatography-tandem mass spectrometry method for the determination of trace concentrations of two of these compounds, JWH-018 and JWH-073, in human blood. Samples underwent liquid-liquid extraction at pH 10.2 into ethyl ether. Tandem mass spectrometry was performed in positive electrospray ionization mode with multiple reaction monitoring using two transitions and one calculated ion transition ratio for each analyte. Deuterated analogs were used as internal standards. Total run time was 2.6 min. The linear dynamic range was 0.05-50 ng/mL with a limit of detection of 0.01 ng/mL for each analyte. Intra-run imprecision (at two different concentration levels, 2 and 8 ng/mL) was 3.9-10.3% for JWH-018 and 3.5-6.2% for JWH-073. Inter-run imprecision was 6.5-7.2% for JWH-018 and 4.8-5.5% for JWH-073. Intra-run accuracy was 95.9-112.7% for JWH-018 and 92.6-104.7% for JWH-073. Inter-run accuracy was 99.1-107.0% for JWH-018 and 97.7-102.0% for JWH-073. Carryover, exogenous drug interferences, ion suppression and matrix selectivity were also assessed. The method has been applied to postmortem forensic casework received by the laboratory and has proven to be robust and reliable. Concentrations of authentic samples have ranged from 0.1-199 ng/mL for JWH-018 and 0.1-68.3 ng/mL for JWH-073.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom