Validation of a Gas Chromatography–Ion Trap-Tandem Mass Spectrometry Assay for the Simultaneous Quantification of Cocaine, Benzoylecgonine, Cocaethylene, Morphine, Codeine, and 6-Acetylmorphine in Aqueous Solution, Blood, and Skeletal Muscle Tissue
Author(s) -
Kelly A. Rees,
Poppy McLaughlin,
M. David Osselton
Publication year - 2012
Publication title -
journal of analytical toxicology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.161
H-Index - 76
eISSN - 1945-2403
pISSN - 0146-4760
DOI - 10.1093/jat/bkr006
Subject(s) - benzoylecgonine , chromatography , chemistry , repeatability , codeine , detection limit , tandem mass spectrometry , solid phase extraction , extraction (chemistry) , morphine , gas chromatography–mass spectrometry , mass spectrometry , quantitative analysis (chemistry) , metabolite , pharmacology , medicine , biochemistry
A gas chromatography-ion trap-tandem mass spectrometry method was developed and validated for the simultaneous solid-phase extraction and quantification of cocaine, benzoylecgonine, cocaethylene, morphine, codeine, and 6-monoacteylmorphine in blood, muscle tissue, and water. An assay for total morphine in blood and muscle was also validated. The limit of quantification was ≤ 0.01 mg/kg in muscle and ≤ 0.005 mg/L in blood and water. Good linearity was observed over the concentration ranges studied (r ≥ 0.99). The repeatability (% RSD) at the three concentration levels was typically ≤ 15% and never exceeded 17%. Intermediate precision of ≤ 16% was obtained for all matrices. Deviation < 20% of the nominal concentration was obtained in all repeatability and intermediate precision experiments. Extraction recoveries for cocaine and metabolites ranged from 91% to 110% in water, 81% to 110% in blood, and 61% to 75% in muscle. Recoveries for opiates ranged from 59% to 104% in water, 50% to 95% in blood, and 41% to 79% in muscle. The hydrolysis efficiency for the total morphine assay in blood and muscle ranged from 91% to 99% with within-day and intermediate precisions of ≤ 14% and ≤ 12%, respectively.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom